• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number L5C4531
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/16/2019
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that two (2) homechoice cassettes had a crack in the patient line which resulted in a leak.This occurred during set up of peritoneal dialysis therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The actual device was not available; however, a companion sample was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.The sample was loaded into an in-house homechoice machine with no issues observed.Functional testing, including leak testing, clear passage testing, and clamp function testing was performed with no issues noted.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
mountain home AR
MDR Report Key9298295
MDR Text Key166163111
Report Number1416980-2019-06206
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00085412090078
UDI-Public(01)00085412090078
Combination Product (y/n)N
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 12/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberL5C4531
Device Lot NumberH19H11025
Was Device Available for Evaluation? No
Date Manufacturer Received12/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
HOMECHOICE DEVICE; UNSPECIFIED PD SOLUTION
-
-