• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 LATEX FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 LATEX FOLEY CATHETER Back to Search Results
Device Problems Leak/Splash (1354); Detachment of Device or Device Component (2907)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 08/01/2019
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the balloon was fixed, and the surface of the catheter leaked fluids resulting in the detachment of the catheter.
 
Manufacturer Narrative
The reported event is inconclusive as no sample was returned for evaluation.A potential root cause for this failure mode could be bad fit with shaft/no drainage eye/ poorly seated balloon /bladder neck interface.The lot number is unknown; therefore, the device history record could not be reviewed.Unable to review the labelling due to unknown product code.Although the product family is unknown, the foley catheter ifus are found to be adequate based on past reviews.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event Description
It was reported that the balloon was fixed, and the surface of the catheter leaked fluids resulting in the detachment of the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATEX FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9298459
MDR Text Key179431444
Report Number1018233-2019-07177
Device Sequence Number1
Product Code EZC
Combination Product (y/n)N
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 11/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Date Manufacturer Received11/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age69 YR
-
-