Model Number 595000-001
Device Problem Audible Prompt/Feedback Problem (4020)
Patient Problem No Patient Involvement (2645)
Event Date 10/24/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the freedom driver was not supporting a patient.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom driver was not supporting a patient.The customer, a syncardia authorized distributor, reported that the freedom driver exhibited a fault alarm during an onboard battery exchange.
 
Manufacturer Narrative
The driver's alarm history was reviewed and did not reveal any recorded alarms.Only permanent fault alarms are recorded in the driver's alarm history; intermittent and/or recoverable alarms, such as battery alarms, are not recorded.The driver in "as received" passed all sections of functional testing.Additionally, an onboard battery exchange test was performed and the driver did not exhibit any alarms.During investigation testing, the customer-reported issue was not reproduced and there was no evidence of a device malfunction.The root cause of the customer-reported fault alarm during an onboard battery exchange could not be conclusively determined.The onboard batteries in use during the customer-reported issue were not returned to syncardia for investigation.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key9298522
MDR Text Key165983160
Report Number3003761017-2019-00323
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 11/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2019
Date Manufacturer Received10/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-