Catalog Number 295600-001
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 10/25/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses low risk to a patient because the issue was observed when the freedom driver was not supporting a patient.In addition, the freedom driver has a redundant power source of multiple onboard batteries.The freedom home ac power supply will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
Tha freedom home ac power supply was not in patient use.The freedom home ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia certified hospital, reported that the freedom home ac power supply was damaged where the cord is inserted and wires were exposed.
 
Manufacturer Narrative
Visual inspection confirmed the customer-reported issue as the bend relief was found to be damaged with exposed wiring and the power brick was damaged at the edges.It cannot be conclusively determined what caused this damage to the bend relief, but it is likely that it was caused by an excessive amount of tension being applied to the power source, toward the plug end.During investigation testing, the power supply met all functional acceptance criteria.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 5035 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM HOME AC POWER SUPPLY
Type of Device
EXTERNAL POWER SUPPLY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key9298524
MDR Text Key165983199
Report Number3003761017-2019-00324
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number295600-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2019
Date Manufacturer Received10/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age56 YR
-
-