Model Number 459878
Device Problems High impedance (1291); Device Contamination with Body Fluid (2317)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/29/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during implant, the cardiac resynchronization therapy defibrillator (crt-d) was missing the setscrew from the left ventricular (lv) port.The crtd was ultimately implanted by removing the grommet from the old device and reusing the setscrew from the old device.The crtd remains in use.It was also noted that left ventricular (lv) lead exhibited high impedance on the distal electrodes due to suspected fluid ingress.The lv lead remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN PERFORMA S MRI SURESCAN
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9306193
MDR Text Key166934923
Report Number2649622-2019-20788
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/25/2020
Device Model Number459878
Device Catalogue Number459878
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/29/2019
Date Device Manufactured06/13/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-