A field service engineering (fse) was at customer's site to resolve reported event.Fse confirmed the reported error by reviewing the error log and reproduced the error by attempting to run a test cup transfer evaluation.The fse cleaned the cup pickup sensor on the hybrid arm and the main arm nozzle which was clogged.Fse also cleaned and calibrated the pressure sensor.Fse validated the analyzer by running quality control and patient samples without errors and within acceptable range.No further action required by field service.The aia-2000 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The aia-2000 operator's manual under appendix 4: error messages states the following: [2151] hybrid arm /cup transfer cup pickup failure.Cause: the cup-gripping sensor failed to detect a cup after the cup pickup operation was performed.Solution: contact tosoh service center or local representatives.The most probable cause of the reported event was due to dirty cup pickup sensor on the hybrid arm, dirty pressure sensor and clogged main arm nozzle.
|