SMITH & NEPHEW ORTHOPAEDICS LTD ACETABULAR CUP HAP SIZE 44/52; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74122152
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Patient Device Interaction Problem (4001)
Patient Problems Pain (1994); Discomfort (2330); Toxicity (2333); Injury (2348); Test Result (2695)
Event Date 08/15/2011
Event Type  Injury  
Event Description
It was reported that a right hip revision surgery was performed due to persistent pain, discomfort, metallosis and elevated test results.
 
Manufacturer Narrative
It was reported that a right hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup and bhr head was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the cup or head.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The available medical documents were reviewed.: based on the information provided, the intraoperative findings of greyish fluid and brownish synovium may be consistent with findings associated with metal debris.However, without the supporting radiographic images, lab/pathology results, and/or the analysis of the explanted components, the reported possibility of aseptic lymphocyte-dominant vasculitis-associated lesion cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the revision is expected post-operative healing/pain.The patient¿s continued right hip/groin pain and two subsequent revisions are not related to s&n devices.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Manufacturer Narrative
New information: g4, d4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETABULAR CUP HAP SIZE 44/52
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key9309497
MDR Text Key165976780
Report Number3005975929-2019-00400
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup,Followup
Report Date 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date04/30/2013
Device Catalogue Number74122152
Device Lot Number08DW16862
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/15/2019
Initial Date FDA Received11/12/2019
Supplement Dates Manufacturer Received06/24/2020
08/18/2020
Supplement Dates FDA Received06/25/2020
10/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
74123144 FEM HEAD 44MM 08HW18442
Patient Outcome(s) Hospitalization; Required Intervention;
-
-