The device was returned to zoll medical corporation and the customer's report was not replicated or confirmed.The device was put through extensive testing including bench handling, defib testing, ecg testing, pace testing and an impedance check without duplicating any malfunction.An internal inspection found no discrepancies.The device was recertified and returned to the customer.Review of the device activity logs showed that a bradycardia rhythm was present even after the device was changed to pacer mode.The configuration used during pacing could not be determined.Instances of ecg lead off were observed in the logs when the user swiched to pacer mode.This suggests that the ecg cable or the one step pacing ecg electrodes were not properly connected on either the device or the patient end.The device was determined to be fully capable of delivery therapy.The multi-function cable, ecg accessory cable, or event electrodes were not returned as part of the investigation.No trend is associated with reports of this type.
|