• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC LAP 5MM SHEAR 36CM; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC LAP 5MM SHEAR 36CM; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HAR36
Device Problems Material Deformation (2976); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/30/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # t92u2d.Investigation summary: the device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.During functional testing on gen11, an alert screen was displayed.A probable cause for the device to stop activating and the gen11 to display an alert screen is blade damage.The device was disassembled to inspect the internal components and no anomalies were found.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damage blade can result in a broken blade.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Event Description
It was reported that during a gastrectomy procedure, the knife of device was deformed by the heat.There were no adverse consequences for the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC LAP 5MM SHEAR 36CM
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key9318988
MDR Text Key176860016
Report Number3005075853-2019-23432
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036014447
UDI-Public10705036014447
Combination Product (y/n)N
PMA/PMN Number
K120729
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 09/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2024
Device Catalogue NumberHAR36
Device Lot NumberT92U2D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2019
Date Manufacturer Received10/18/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-