• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER (FOR INSERTION VIA SUPERIOR VENA CAVA); CATHETER, FLOW DIRECTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER (FOR INSERTION VIA SUPERIOR VENA CAVA); CATHETER, FLOW DIRECTED Back to Search Results
Model Number D97120F5
Device Problem Failure to Capture (1081)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/01/2019
Event Type  malfunction  
Manufacturer Narrative
Without the return of the product, it is not possible to determine if damages or defects exist on the product, nor can any manufacturing nonconformance, failure mode, root cause, or potential contributing factors be identified.A device history record review was completed and documented that the device met all specifications upon distribution.An engineering evaluation has been initiated to assess for any manufacturing-related processes which could be correlated to the complaint.Swan-ganz pacing thermodilution (td) catheters serve as diagnostic and therapeutic tools in the management of critically ill patients.There are multiple failure modes that may require the exchange of a pacing catheter.Since proper functioning of the pacing catheter depends on the electrical continuity of its electrodes and internal wires, care should be exercised when handling the catheter.Stretching, kinking, or forceful wiping of the catheter may result in damage.After stable pacing has been confirmed, the proximal end of the catheter should be secured to the insertion site to prevent undue movement that could result in tip dislodgment and loss of capture, or catheter migration.Care should be taken not to kink the catheter body when securing it.In this complaint, it could not be determined if procedural factors or device handling may have contributed to the reported event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.Udi # (b)(4).
 
Event Description
It was reported that a swan-ganz pacing catheter from lot 62451093 failed to capture during use in the or.There was on patient injury.Patient demographics were requested and not provided.The catheter was disposed of before the customer could retrieve it for return evaluation, as it was ¿a busy room with an emergent situation going on.¿.
 
Manufacturer Narrative
Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER (FOR INSERTION VIA SUPERIOR VENA CAVA)
Type of Device
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
MDR Report Key9321895
MDR Text Key194617116
Report Number2015691-2019-04219
Device Sequence Number1
Product Code DYG
Combination Product (y/n)N
PMA/PMN Number
K822723
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 11/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Model NumberD97120F5
Device Catalogue NumberD97120F5
Device Lot Number62451093
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/01/2019
Initial Date FDA Received11/14/2019
Supplement Dates Manufacturer Received07/23/2020
Supplement Dates FDA Received01/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-