NOVO SURGICAL, INC. KIRSCHNER TRACTION BOW; COMPONENT, TRACTION, NON-INVASIVE Back to Search Results
Device Problem Electromagnetic Interference (1194)
Patient Problem Pain (1994)
Event Date 11/07/2019
Event Type  Injury  
Event Description
Patient went into 1.5 mri unit with kirschner bow used for skeleton traction.It is a newer product being used within the hospital, the techs on the evening of the event did not check, they thought it was just another fixator.When the patient got close to the magnet, the patient starts to complain of pain, patient was removed from magnet.This is when the techs discovered it was ferrous.The patient was sent back to his room, had the traction device removed and rescanned without incident.No further pain or harm to patient.Fda safety report id # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIRSCHNER TRACTION BOW
Type of Device
COMPONENT, TRACTION, NON-INVASIVE
Manufacturer (Section D)
NOVO SURGICAL, INC.
MDR Report Key9321961
MDR Text Key166737458
Report NumberMW5091077
Device Sequence Number1
Product Code KQZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/13/2019
Was Device Evaluated by Manufacturer? No Information
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age21 YR