Catalog Number H1-M |
Device Problems
Difficult to Remove (1528); Detachment of Device or Device Component (2907); Physical Resistance/Sticking (4012)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/15/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Physician intended to use a hawkone device during treatment of a lesion in the patient¿s superficial femoral artery (sfa).Ifu was followed.It is reported that the tip broke off two hawkone devices.No patient injury reported.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: the lesion was calcified.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: there was resistance felt when advancing the first hawkone device.Excessive force was not used.It was reported that the tip separated at the hinge pin and the tip of the hawkone detached within the patient during advancement of the device.The tip detached within the target lesion.The tip did not embolize during the procedure.A snare was used to retrieve the tip.The second hawkone was then opened and resistance was felt during advancement.Excessive force was used.The tip separated at the hinge pin during device withdrawal inside the vessel, but the tip did not detach within the patient.The tip did not embolize during the procedure.The procedure was completed with a balloon angioplasty and reported to be then closed up.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|