(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g4, g7, h1, h2, h3, h6, h10.The specified identity of the device was confirmed and the device was examined at zimmer biomet westminster colorado spine facility.Visual inspection revealed that the probe is bent on both ends and the ball on one end has fractured off.Review of the device history record identified no deviations or anomalies during receipt of product.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|