This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b1, b4, b5, d4, d10, g4, g7, h2, h3, h6 and h10.The device history record (dhr) and previous repair report reviewed noted no related non-conformances, requests for deviation, change notices or any other issues with manufacturing or with the device.The dhr and previous repair report review also found that all verifications, inspections and tests were successfully completed.Product review of the air dermatome by zimmer biomet taiwan on november 4, 2019 revealed that the motor speed was below specifications.The calibration was out of specifications at all settings and the control bar was in the correct position.The customer width plates were not returned for evaluation.Repair of the air dermatome was performed by zimmer biomet taiwan on november 6, 2019 which included replacement of the motor, bearings, o-ring, seal, reciprocating arm, external e-ring, machined head, and throttle hinge gasket.Air dermatome, serial number (b)(4), was then tested and functioned properly.It was repaired, inspected and tested.While the returned product investigation confirmed that the air dermatome was out of calibration at all settings, it cannot be determined from the information provided what actually caused the reported event.Therefore, based on the information provided, a specific root cause of the reported event cannot be determined.The device was noted to be functioning as intended after the unit was recalibrated and the motor, bearings, o-ring, seal, reciprocating arm, external e-ring, machined head, and throttle hinge gasket were replaced.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
|