Model Number PARADYM RF SONR CRT-D 9770 |
Device Problems
High impedance (1291); Battery Problem (2885); Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
Ventricular Fibrillation (2130)
|
Event Date 10/09/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device model involved in this mdr is not approved in the united states; however, it is similar to a device manufactured by microport crm that was cleared or approved by fda for marketing in the united states.
|
|
Event Description
|
On (b)(6) 2019, the subject device in was found in end of service (eos), with a battery voltage of 1.78v and with high impedance in both coils (>3000ohms).The last charge time was 40s.The device was re-interrogated on (b)(6) 2019.High impedances were still observed.
|
|
Event Description
|
On october 8th 2019, the subject device in was found in end of service (eos), with a battery voltage of 1.78v and with high impedance in both coils (>3000ohms).The last charge time was 40s.The device was re-interrogated on (b)(6) 2019.High impedances were still observed.On (b)(6) 2019, the subject device was explanted and replaced.The right ventricular lead was abandoned and replaced.
|
|