Model Number X SERIES |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during biomed testing, the device was unable to detect the attached multi-function cable.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation; the reported malfunction was observed and attributed to a system interconnect flex cable that was not properly seated.The cable was reseated to correct the reported problem.Analysis for reports of this type has not identified an increase in trend.
|
|