Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the returned instrument noted that there were no abnormalities that would have caused or contributed to the reported condition.The returned instrument was found to function properly and the test needle remained intact and attach to the instrument throughout the testing process.No difficulty was experienced in loading, unloading or toggling the test needle in the returned instrument.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
According to the reporter, during a laparoscopic sleeve gastrectomy procedure, the device was difficult to toggle.It was also reported that the needle fell out of the instrument and was retrieved with a needle holder.The needle was removed from the patient's cavity and placed on the needle count sheet making sure that all pieces were located.A new loading unit and instrument were used to complete the procedure.There was no patient injury.
|