Model Number 0279401200 |
Device Problems
Self-Activation or Keying (1557); Failure to Shut Off (2939)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/06/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
|
Event Description
|
It was reported that the device was continuously activating.
|
|
Manufacturer Narrative
|
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: likely functioning automatically as the probe button and the footswitch was not pressed.The probable root cause/s could be probe short (due to fluid ingress), button stuck on firing position.
|
|
Event Description
|
It was reported that the device was continuously activating.
|
|