The reported event was confirmed.Visual inspection noted one bulb syringe was received.Visual evaluation noted a piece of hair measuring 0.2730" embedded in the shaft of the syringe.This is out of specification per inspection procedure which states, "loose or embedded foreign matter greater than 0.6mm2 or 1/16¿ in length is not permitted.(3 particles maximum per side or surface).Per tappi dirt estimation chart." although the reported event was confirmed, the root cause could not be determined.A potential root cause for this failure could be, ¿no follow up to the production areas cleaning procedure.¿ the lot number is unknown; therefore, the device history record could not be reviewed.A labeling review was not performed because the user alleged the presence of foreign material on the product, which the user did not cause.Labeling does not apply in this case.
|