The reported event was confirmed, however, the cause is unknown.Visual evaluation of the sample noted one used silicone foley with sample port connector and cut drainage tubing within opened original packaging.The catheter balloon was inflated with 10 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) and immediately leaked from a (0.4655") tear in balloon surface.There were no missing balloon pieces.The balloon must not be torn per the standard.Although the reported event was confirmed, the root cause could not be determined.A potential root cause for this failure could be tooling damage (core pins).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "[warnings].1.Method for use (1) do not inflate the balloon in the urethra.[the urethra may be injured.] (2) do not pull the catheter hard.[the bladder/urethra may be injured.] 2.Applicable patients.Patients with delirium who might pull out catheter [when patient tugs at catheter unconsciously, the bladder and urethra may be damaged.] [contraindications].1.Method for use: (1) do not reuse (2) do not resterilize.(3) this device contains 10% povidine-iodine.For patients with past history of allergic hypersensitivity to povidone-iodine or iodine, consider using alternative disinfectants.(4) be careful that the catheter is not exposed to ointments, contract medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (5) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] directions for use: 1) clean the area around the external urethral meatus with the packaged cotton balls immersed in the antiseptics.2) lubricate the catheter shaft with the lubricant jelly.3) carefully insert the catheter into the urethral meatus.After the balloon advanced in the bladder, attach the needleless syringe, and gently infuse the specified volume of sterile water to inflate the balloon.4) pull the catheter slightly to seat the balloon at the level of the bladder neck.5) to deflate and remove the balloon, attach a needleless syringe to let sterile water in the balloon come out spontaneously through balloon deflation without aspiration.After balloon deflation, withdraw the catheter slowly while confirming that no abnormal resistance is encountered.".
|