ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER EDGEPLUS VALVED ENTRY SYSTEM; CANNULA, TROCAR, OPHTHALMIC
|
Back to Search Results |
|
Catalog Number 8065751585 |
Device Problem
Dull, Blunt (2407)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/23/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported that two of the trocars would not go into the eye as if they were too dull.The surgeon used another product and completed the procedure.There was no patient harm.
|
|
Manufacturer Narrative
|
Additional information provided in d.10, h.3, h.6 and h.10.Ten unopened trocar assemblies were received in pouches for evaluation.The samples were visually inspected and were found to be conforming.The samples were then functionally tested for penetration and were found to be conforming.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The returned unopened samples were found to be visually and functionally conforming, therefore dull trocar blades as described in the complaint were not confirmed.However, since the actual complaint samples were not returned, a root cause cannot be determined for the complaint as described by the customer.Because the actual complaint samples were not returned the exact root cause for this complaint is unknown, specific action with regards to this complaint cannot be taken.All trocar knives are 100% inspected by trained operators using a minimum of 10x magnification during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No further actions are required.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|
|
|