STANMORE IMPLANTS WORLDWIDE JTS DRIVE UNIT; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Catalog Number UNK_STM |
Device Problems
Use of Device Problem (1670); Insufficient Information (3190)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); Cancer (3262)
|
Event Date 09/06/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Please note that this event is related to a jts drive unit which is used in conjunction with the jts non-invasive extendible implant (k092138).An investigation is being performed in an attempt to identify the cause of the event.Should additional information become available it will be reported in a supplemental report.
|
|
Event Description
|
This pi is created for the first attempt last 25 of august.It is reported " first attempt has been performed with no success last 25 of august with power unit set on 0.Second attempt today." additional notes state "center is exposed to jts system, never got problems before.Same skilled physician as in the past.It has been tried several time the unlocking procedure to switch from b to a.Power unit set on 3 and 4.Implant sounded as locked, sounded working going reverse (b) and then going ahead (a) but just for a few seconds.".
|
|
Event Description
|
This pi is created for the first attempt last 25 of august.It is reported " first attempt has been performed with no success last 25 of august with power unit set on 0.Second attempt today." additional notes state "center is exposed to jts system, never got problems before.Same skilled physician as in the past.It has been tried several time the unlocking procedure to switch from b to a.Power unit set on 3 and 4.Implant sounded as locked, sounded working going reverse (b) and then going ahead (a) but just for a few seconds.".
|
|
Manufacturer Narrative
|
On initial receipt of the reported event, the event description was reviewed and with the limited information available at that time, a decision was made to report the event.The investigation confirmed that this unit was functional and did not cause or contribute to the reported event.The event is therefore, not reportable.
|
|