DEPUY FRANCE SAS - 3003895575 DXTEND SCREW LOCK D4.5X24MM; DELTA XTEND IMPLANTS : SCREWS Back to Search Results
Catalog Number 130790024
Device Problems Loss of or Failure to Bond (1068); Osseointegration Problem (3003)
Patient Problems Inadequate Osseointegration (2646); No Code Available (3191)
Event Date 10/25/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The patient was revised to address loosening of the unknown component at the bone to implant interface.It was also reported that the patient had a well functioning reverse shoulder, when he fell and dislodged the glenoid component, breaking 2 screws in the process.The entire construct (metaglene, glenosphere, 4 screws, was loose due to the trauma and came out in one piece.) surgeon revised it to a tornier screw in baseplate, but left our stem in and replaced a 42 poly.Two of the screws were broken (42/36, and the broken pieces were left in the glenoid.Doi: (b)(6) 2016; dor: (b)(6) 2019; right shoulder.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND SCREW LOCK D4.5X24MM
Type of Device
DELTA XTEND IMPLANTS : SCREWS
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
Manufacturer (Section G)
DEPUY FRANCE SAS 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR   69801
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester, PA 19380-0988
6103142063
MDR Report Key9355529
MDR Text Key167784660
Report Number1818910-2019-117406
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295027836
UDI-Public10603295027836
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120174
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number130790024
Device Lot Number5274374
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/18/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight92
-
-