|
Model Number FGS-0636 |
Device Problem
Loss of or Failure to Bond (1068)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
This product is not sold in us and is 510(k) exempt.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, they had a capsule which failed to attach.There was no harm to the patient, no intervention was required, and a repeat procedure was performed.There was nothing unusual about the patient or the procedure and an endoscopy had been performed prior to the procedure and showed the esophagus to be normal.A lubricant was used to facilitate placement of the capsule and the delivery system and capsule will be returned for investigation.
|
|
Manufacturer Narrative
|
Evaluation summary: this report is based on information provided by the complaint tracking system.A product sample was provided by customer to medtronic cs at (b)(4) site unit but was not provided to (b)(4) investigation team.Per tech support description there is no enough information to determine if product whether or not product meet spec.Since no sample arrived at (b)(4) for investigation, visual inspection cannot be performed.The customer reported they had a capsule which failed to attach.Per tech support description there is no enough information to determine if product whether failure has been confirmed or not.The investigation performed by tech support did not determine the cause of the reported issue.A review of the device history records indicated that this serial/lot number was released meeting all specifications as manufactured.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: first name, last name, phone number, email.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, they had a capsule which failed to attach.There was no harm to the patient, no intervention was required, and a repeat procedure was performed.There was nothing unusual about the patient or the procedure and an endoscopy had been performed prior to the procedure and showed the esophagus to be normal.A lubricant was used to facilitate placement of the capsule and the delivery system and capsule will be returned for investigation.
|
|
Manufacturer Narrative
|
Additional information: g4, h3, h6.H3 evaluation summary: this report is based on information provided by medtronic investigation personnel and bravo device arriving in bio lab.One bravo device was received for evaluation.The returned sample met specification as received by medtronic.The customer reported they had a capsule which failed to attach.The reported condition was not confirmed.The plunger was fully released (6 ribs), the vacuum reached 510 mmhg and was applied for at least 30 seconds.The wire that holds the capsule was completely off and the foam gasket was in good condition.The investigation found the device to function normally and within specifications.A review of the device history records indicated that this serial/lot number was released meeting all specifications as manufactured.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|