• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Display Difficult to Read (1181); Display or Visual Feedback Problem (1184)
Patient Problem No Patient Involvement (2645)
Event Date 10/28/2019
Event Type  malfunction  
Manufacturer Narrative
On (b)(6) 2019, during preventive maintenance, the lcd screen back-light of the autopulse platform (sn (b)(4)) was flickering and the screen was difficult to read.The root cause for the observed lcd problem was due to the defective lcd display as a result of aging of the device.The autopulse platform was manufactured in 2012 and is 7 years old, past the expected service life of 5 years.The lcd display was replaced to remedy the problem.Upon visual inspection, noticed flickering lcd screen.No physical damage was observed.During functional testing, the autopulse platform performed compressions without any fault or error.The archive data could not be downloaded due to the defective processor pca (printed circuit assembly).The processor pca was replaced to remedy the problem.Following service, the autopulse was subjected to the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test batteries until discharged without any fault or error.The brake gap inspection was performed and verified the brake gap was within the specification.Load cell characterization test was performed and confirmed both cell modules are functioning within the specification.The autopulse platform passed the final testing without any fault or error.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaint reported for autopulse platform with sn (b)(4).
 
Event Description
On (b)(6) 2019, during preventive maintenance, the lcd screen back-light of the autopulse platform (sn (b)(4)) was flickering and the screen was difficult to read.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key9363292
MDR Text Key170106569
Report Number3010617000-2019-01064
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111001274
UDI-Public00849111001274
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0740-01-66
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/28/2019
Initial Date FDA Received11/22/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-