• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN KNEE FEMORAL ADAPTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN KNEE FEMORAL ADAPTOR Back to Search Results
Catalog Number UNK KNEE FEMORAL ADAPTOR
Device Problems Corroded (1131); Degraded (1153)
Patient Problems Edema (1820); Pain (1994); Skin Irritation (2076); Test Result (2695); No Code Available (3191)
Event Date 04/04/2012
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
The literature article entitled, "adverse local tissue response lesion of the knee associated with morse taper corrosion" written by william c.Mcmaster, md and jay patel, md published by the journal of arthroplasty vol.28 no.2 2013 accepted by publisher 4 april 2012 was reviewed.The article's purpose was to report on a case report of a (b)(6) year old man with a left tka who ten received a depuy lps in 2009 by revision to original implant (product not specified).Approximately 20 months post lps implantation he begin to experience swelling, loss of motion, and pain.He developed local knee skin rash with redness.Co levels were elevated at 30 mcg/l and pain progressed.Tests were negative for infection and no radiographic findings for cause of pain.Surgical exploration revealed black encrustations at the morse taper junction of the distal femoral component and well ingrown porous-coated intramedullary femoral stem extension and no signs of wear on the poly insert or bearing surfaces.Strong force required to disengage the taper junction where the implants presented with corrosion on both sides.Capsular debridement was performed but no significant signs of alval were present.A new distal femoral component was applied to the original ingrown femoral stem as a temporary fix and the hinged components were then reassembled.Patient healed without further complications.The article does not identify cement manufacturer, and no mention of patella resurfacing.Depuy products utilized: hinged lps with rotating platform.Adverse events: pain, swelling, skin reaction, joint loss of motion (attributed to pain), blood heavy metal increased, surgical intervention, implant corrosion at femoral adaptor/sleeve component junction.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: corrected: adverse event problem (device code).Product complaint # (b)(4).Investigation summary: the device associated with the reported event was not returned to the investigation site for evaluation.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN KNEE FEMORAL ADAPTOR
Type of Device
KNEE FEMORAL ADAPTOR
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9364394
MDR Text Key185085028
Report Number1818910-2019-114992
Device Sequence Number1
Product Code JWH
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type literature
Type of Report Initial,Followup
Report Date 11/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK KNEE FEMORAL ADAPTOR
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/01/2019
Initial Date FDA Received11/22/2019
Supplement Dates Manufacturer Received12/30/2019
Supplement Dates FDA Received12/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-