A field service engineer (fse) was at the customer site to address the reported issue.The fse was not able to reproduce the issue.The fse replaced the specimen dispense syringe and resolved the issue.Instrument validation was performed by quality control (qc).Qc results passed within the published ranges.The aia-900 analyzer returned to operation.The fse returned to the customer site several days after where the customer reported bhcg result on the aia-900 analyzer was <0.5 miu/ml compared to lab core methodology.The fse could not reproduce the issue.The fse verified and adjusted all sample probe tip bottoms, liquid sense, and level detection to resolve the issue.Instrument validation was performed by quality control (qc).Qc results passed within the published ranges.The aia-900 analyzer returned to operation.No further action required by field service.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 23sep2018 to aware date (b)(6) 2019.One other similar complaint was identified during the search period.The st aia-pack st bhcg analyte application manual states the following: limitations of the procedure.For diagnostic purposes, the results obtained from this assay should be used in conjunction with other data (e.G., symptoms, results of other tests, clinical impressions, therapy, etc.).Using st aia-pack bhcg, the highest concentration of total beta hcg measurable without dilution is 400 miu/ml, and the lowest measurable concentration is 0.5 miu/ml (assay sensitivity).Although the approximate value of the highest calibrator is 200 miu/ml, the exact concentration may be slightly different.The assay specification, assay range high, should be defined as the upper limit of the assay range, 400 miu/ml.Although hemolysis has an insignificant effect on the assay, hemolyzed samples may indicate mistreatment of a specimen prior to assay and results should be interpreted with caution.Lipemia has an insignificant effect on the assay except in the case of gross lipemia where spatial interference may occur.Specimens from patients who have received preparations of mouse monoclonal antibodies for diagnosis or therapy may contain human anti-mouse antibodies (hama).Such specimens may show falsely elevated values when tested for total beta hcg.Certain medications may interfere with assay performance.Specimens from patients taking medicines and/or medical treatment may show erroneous results.All results should be interpreted with respect to the clinical picture of the patient.For a more complete understanding of the limitations of this procedure, please refer to the specimen collection and handling, warnings and precautions, storage and stability, and procedural notes sections in this insert sheet.Expected values.Each laboratory should determine a reference interval which corresponds to the characteristics of the population being tested.As with all diagnostic procedures, clinical results must be interpreted with regard to concomitant medications administered to the patient.The probable cause of the issue is pending investigation.
|