An event of the device having an unstable position, the defect being larger after the occluder was deployed and removed, hypotension and tachycardia was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
On (b)(6) 2019, a 12 mm amplatzer muscular vsd occluder was selected for a postinfarct vsd closure.The device was deployed in the vsd, but due to an unstable position it was recaptured/redeployed; it was aligned horizontally and not in-line with the septum.After recapturing the device, the defect was larger than prior to the procedure.The patient became hypotensive and tachycardic, ecmo therapy was initiated, and the patient was sent to cardiac surgery for vsd closure.The defect was successfully closed with a patch.The patient is in the icu and remains on ecmo support.The user indicated that this was a complicated surgery due to the unstable tissue (post-ischemic, post-infarct) of the defect.
|