Model Number 29MJ-501
Device Problem Material Separation (1562)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/12/2019
Event Type  Injury  
Event Description
On (b)(6) 2019, a 29 mm masters series mechanical heart valve was selected for implant.While rotating the valve, a leaflet dislodged.The leaflet and valve were removed and the device was exchanged with another masters valve.The patient remained hemodynamically stable, but it took an hour to arrange the replacement valve resulting in a clinically significant delay in the procedure.
 
Manufacturer Narrative
The reported event of a dislodged leaflet was confirmed.One leaflet was dislodged from the orifice and fractured.The recessed pivot area was also chipped.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the leaflet dislodgment and fracture could not be conclusively determined.
 
Manufacturer Narrative
An event of a dislodged leaflet was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.Based on the information received, the cause of the reported incident could not be conclusively determined.Please note, per the instructions for use artmt60008090 rev.A, "using the valve holder/rotator and an sjm¿ valve holder handle, rotate the valve in situ to the desired position.The valve should rotate freely.If resistance is noted, the valve holder/rotator may not be properly seated in the valve, the valve may not be in the fully closed position, or the valve may be oversized.If the valve does not rotate freely, do not force valve rotation.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES MECHANICAL HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key9368733
MDR Text Key167781736
Report Number2648612-2019-00104
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734006552
UDI-Public05414734006552
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 08/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/25/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/20/2023
Device Model Number29MJ-501
Device Catalogue Number29MJ-501
Device Lot Number6701918
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2020
Was the Report Sent to FDA? No
Date Manufacturer Received08/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight67
-
-