On (b)(6) 2019, a 29 mm masters series mechanical heart valve was selected for implant.While rotating the valve, a leaflet dislodged.The leaflet and valve were removed and the device was exchanged with another masters valve.The patient remained hemodynamically stable, but it took an hour to arrange the replacement valve resulting in a clinically significant delay in the procedure.
|
An event of a dislodged leaflet was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.Based on the information received, the cause of the reported incident could not be conclusively determined.Please note, per the instructions for use artmt60008090 rev.A, "using the valve holder/rotator and an sjm¿ valve holder handle, rotate the valve in situ to the desired position.The valve should rotate freely.If resistance is noted, the valve holder/rotator may not be properly seated in the valve, the valve may not be in the fully closed position, or the valve may be oversized.If the valve does not rotate freely, do not force valve rotation.".
|