• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD ENDO STITCH; ENDOSCOPIC TISSUE APPROXIMATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD ENDO STITCH; ENDOSCOPIC TISSUE APPROXIMATION DEVICE Back to Search Results
Model Number 173016
Device Problems Break (1069); Component or Accessory Incompatibility (2897); Detachment of Device or Device Component (2907)
Patient Problems Radiation Exposure, Unintended (3164); No Code Available (3191)
Event Date 11/07/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a total laparoscopic hysterectomy, while closure of vaginal cuff, the operator had trouble loading the needle onto the handle.It was also noted that the needle on device broke in half and fell into the patient's abdominal cavity.Needle was found with x-ray near the intestines.
 
Manufacturer Narrative
Additional information: d10, g4, h3, h6 h3 evaluation summary: post market vigilance (pmv) led an evaluation of three devices.Visual inspection of the returned products noted that there were no abnormalities that would have caused or contributed to the reported condition.The returned instruments were loaded with test needles from the post marketing vigilance (pmv) inventory and applied to test media for functional testing.The returned instruments were found to function properly and test needles remained intact.No difficulty was experienced in loading, unloading or toggling the test needles in the returned instruments.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO STITCH
Type of Device
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key9371500
MDR Text Key167868855
Report Number9612501-2019-02265
Device Sequence Number1
Product Code OCW
UDI-Device Identifier10884521100008
UDI-Public10884521100008
Combination Product (y/n)N
PMA/PMN Number
K934738
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Model Number173016
Device Catalogue Number173016
Device Lot NumberJ8L2562EY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2019
Initial Date Manufacturer Received 11/07/2019
Initial Date FDA Received11/25/2019
Supplement Dates Manufacturer Received02/06/2020
Supplement Dates FDA Received02/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age31 YR
Patient Weight63
-
-