• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/07/2019
Event Type  malfunction  
Manufacturer Narrative
The sheath, 4fc12 with lot number 83972, was returned and analyzed.Visual inspection of the sheath showed it was intact with no apparent issues.Air aspiration was reproduced during aspiration of the sheath with a balloon test catheter introduced.In conclusion, the reported air ingress was confirmed through testing.The sheath failed the returned product inspection due to the air ingress during aspiration.If information is provided in the future, a supplemental report will be issued.
 
Event Description
After a completed case, the sheath subsequently tested out of specification per the manufacturer's investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9372389
MDR Text Key195104891
Report Number3002648230-2019-00805
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/25/2021
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number83972
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2019
Date Manufacturer Received11/05/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/26/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-