MEDTRONIC, INC. VIVA¿ XT CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTBA2D1
Device Problems Failure to Deliver Shock/Stimulation (1133); Capturing Problem (2891)
Patient Problem Ventricular Tachycardia (2132)
Event Date 01/01/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
A journal article was reviewed that contained information regarding a cardiac resynchronization therapy defibrillator (crt-d).It was reported that the patient experienced ventricular tachycardia (vt).When a test pace would occur, it was interrupted by ventricular sensing and pacing due to prolonged atrioventricular intervals.The capture threshold test was aborted due to sustained monomorphic vt.It was initially thought that the capture management test was causing vt since it postponed therapy for a relatively long time, resulting in the vt, however, it was determined that capture management was functioning as programmed.The device remains inuse.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA¿ XT CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9373972
MDR Text Key168278468
Report Number2182208-2019-02118
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Reporter Occupation Physician
Type of Report Initial
Report Date 11/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date02/28/2019
Device Model NumberDTBA2D1
Device Catalogue NumberDTBA2D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/23/2019
Initial Date FDA Received11/26/2019
Date Device Manufactured08/29/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age58 YR
-
-