• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AESTIVA 7900; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AESTIVA 7900; ANESTHESIA GAS MACHINE Back to Search Results
Device Problems Detachment of Device or Device Component (2907); Infusion or Flow Problem (2964)
Patient Problem No Patient Involvement (2645)
Event Date 11/04/2019
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the equipment and confirmed the reported complaint.The adjustable pressure limiting valve was replaced, and the unit was returned to service.No report of patient involvement.
 
Event Description
The hospital reported the adjustable pressure limiting knob popped off, preventing manual ventilation.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESTIVA 7900
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
john szalinski
MDR Report Key9375810
MDR Text Key217728686
Report Number2112667-2019-02373
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K023366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 11/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/04/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-