According to the reporter, during a laparoscopic sigmoidectomy, at the second firing, the handle was felt stiff, so it was stopped using.When fired, at the third firing, the product clamped nothing, the counter indication suddenly turned to 0, and the jaws would not open.The procedure was completed with another device.There was no patient injury.
|
Additional information: d10, g4, h3, h6 h3 evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the instrument noted the handle was flaccid and not attached to the internal components.It was also noted that the clip counter was no longer active.The instrument was dismantled for visualization of internal components.It was noted that the counter arm was properly seated within handle but that the wishbone link which attaches the trigger to the firing mechanism had disengaged.The wishbone link was reattached to the trigger handle, the counter was removed for further analysis, and the instrument was reassembled.The instrument was then found to cycle without binding.Fifteen clips loaded into the jaws, formed properly, released from the jaws and remained securely attached to test media.When the cartridge was empty, the interlock engaged to prevent the jaws from approximating.A pmv representative battery was assembled onto the subject counter and the counter was manually indexed from 16 to 00 without issues multiple times.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the disengaged wishbone link condition may occur if the handle is forcefully pulled open prior to fully completing the full handle compression.The root cause of the observed damage was due to the product not being used as indicated which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|