Model Number 3820 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/07/2019 |
Event Type
malfunction
|
Event Description
|
It was reported that the delivery shaft was fractured.The target lesion was in the calcified left coronary artery.A 06/3.00 flextome cutting balloon was selected for use.During the procedure, it was noted that the delivery shaft was fractured inside patient.The device was then directly and completely removed from the patient's body without any intervention.The procedure was completed with another of the same device.No complications were reported and patient was stable.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: the device was returned for analysis.The device was received in two sections as a result of a complete breaks of the hypotube.A visual and tactile examination a complete break in the hypotube of the device.The break was located at approximately 415mm distal of the strain relief.The hypotube was also noted to be kinked at more than one location.This type of damage is consistent with excessive force being applied to the delivery system.A visual examination identified that the balloon was not subjected to positive pressure.An examination of the balloon identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination identified no damage to the tip, markerbands or blades of the device.All blades were present and fully bonded to the balloon material.No other issues were identified during the product analysis.
|
|
Event Description
|
It was reported that the delivery shaft was fractured.The target lesion was in the calcified left coronary artery.A 06/3.00 flextome cutting balloon was selected for use.During the procedure, it was noted that the delivery shaft was fractured inside patient.The device was then directly and completely removed from the patient's body without any intervention.The procedure was completed with another of the same device.No complications were reported and patient was stable.
|
|
Search Alerts/Recalls
|