Model Number 45007
Device Problems Entrapment of Device (1212); Device Damaged by Another Device (2915)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/01/2019
Event Type  malfunction  
Event Description
It was reported that the device became stuck on the wire.A 2.1mm jetstream xc catheter and non-bsc filter guidewire were selected for use in an atherectomy procedure.Near the end of treatment in the first treatment location, the device became stuck on the wire.The physician was able to remove the device from the wire but noted that the wire was sticky which prevented the device from moving smoothly.It was thought that the coating of the wire may have come off, causing the device to become stuck.The wire was removed and replaced with a new wire.A new jetstream device, same model, was selected for use to resume treatment.The procedure was finished without any complications, and no adverse patient effects were reported.The procedure ended well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM XC
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road cork
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9378244
MDR Text Key173943385
Report Number2134265-2019-14463
Device Sequence Number1
Product Code MCW
UDI-Device Identifier08714729889892
UDI-Public08714729889892
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number45007
Device Catalogue Number45007
Device Lot Number0024426363
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/06/2019
Initial Date FDA Received11/26/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/13/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-