Model Number R SERIES |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device displayed "ecg disabled" and "ecg fault 200" error messages.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The customer's report was duplicated and the analog board was replaced to remedy the reported problem.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
|
Search Alerts/Recalls
|