• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AISYS; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AISYS; ANESTHESIA GAS MACHINE Back to Search Results
Model Number 1011-9000-000
Device Problems Detachment of Device or Device Component (2907); Device Fell (4014)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Ge healthcare's investigation into the reported occurrence is still ongoing.A follow-up report will be issued when the investigation has been completed.No report of patient involvement.Incident date not provided.Device evaluation anticipated, but not yet begun.
 
Event Description
The hospital reported that a caster was stopped at the base causing the unit to tip.There was no report of patient involvement.
 
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The caster base was ordered for replacement to resolve the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
MDR Report Key9383411
MDR Text Key219764517
Report Number2112667-2019-02380
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K042154
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 01/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1011-9000-000
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-