A field service engineering (fse) was at customer's site to resolve reported event.The fse was able to confirm complaint by examining the linearity and qc results but was unable to reproduce the problem.Fse resolved the reported issue by adjusting the incubator and detector temperatures.The analyzer was validated by running fsh calibration, linearity and qc; all results passed without error and within acceptable range.No further action required by field service.The aia-900 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no similar complaints identified during the search period.The st aia-pack fsh analyte application manual states the following: procedural notes: if a specimen follicle-stimulating hormone concentration is found to be greater than the linearity limit of the assay, 200 miu/ml; the specimen should be diluted with the aia-pack fsh sample diluting solution and re-assayed according to the assay procedure.The recommended dilution for samples containing greater than 200 miu/ml is 1:10 or 1:100.It is desirable to dilute the sample so that the diluted sample reads between 5 and 200 miu/ml.The dilution factor should be entered into the software.For further information on the dilution of specimens, refer to the aia system operator's manual.Evaluation of results quality control in order to monitor and evaluate the precision of the analytical performance, it is recommended that commercially available control samples be assayed daily.The minimum recommendations for the frequency of running internal control material are: after calibration, two levels of controls are run in order to accept the calibration curve.The two levels of controls are also repeated after calibration when certain service procedures are performed (e.G.Temperature adjustment, sampling mechanism changes, maintenance of the wash probe or detector lamp adjustment or change).After daily maintenance, at least two levels of the control should be run in order to verify the overall performance of the tosoh aia system analyzers.If one or more control sample value(s) is out of the acceptable range, it will be necessary to investigate the validity of the calibration curve before reporting patient results.Standard laboratory procedures should be followed in accordance with the regulatory agency under which the laboratory operates.The most probable cause of the reported event was due to incubator & detector temperature adjustment.
|