Reportable based on device analysis completed on 13nov2019.It was reported that the balloon was damaged.The severely stenosed target lesion was located in the severely tortuous and severely calcified left anterior descending artery.During the procedure, the balloon was delivered to the distal end of the lesion by guidezilla guide extension catheter; however, it was noted that the balloon could not be filled at 6atm.Furthermore, it was found that the balloon was damaged.The device was completely removed from the patient without any intervention and the procedure was completed with another of the same device.There were no complications reported and the patient's condition was stable post procedure.However, device analysis revealed a balloon hole/tear detected 2mm proximal of the proximal markerbands.
|