• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA CANADA CORP. PERCEVAL SUTURELESS AORTIC HEART VALVE; TISSUE HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA CANADA CORP. PERCEVAL SUTURELESS AORTIC HEART VALVE; TISSUE HEART VALVE Back to Search Results
Model Number PVS25
Device Problem Patient-Device Incompatibility (2682)
Patient Problem Complete Heart Block (2627)
Event Date 09/06/2019
Event Type  Injury  
Event Description
On (b)(6) 2019 a patient received a perceval pvs25 as part of the sure-avr trial.The manufacturer was notified that on (b)(6) 2019 the patient received a pacemaker implant for a avb iii.
 
Manufacturer Narrative
The manufacturing and material records for the perceval heart valve, model#: icv1210, s/n#: (b)(6), as they pertain to the reported event, were retrieved and reviewed by quality engineering at livanova canada corp.The results confirmed that this valve satisfied all material, visual, and performance standards required for a (model#: icv1210) perceval heart valve at the time of manufacture and release.As the device was not received for analysis, no device investigation can be performed and the root cause of the event cannot be determined.However, based on the document review performed, no manufacturing deficiencies were identified.Conduction disorders are common in patients with aortic valve diseases.As a result, it is common for patients to receive permanent pacemaker implantation after aortic valve replacement.Risk factors include preexisting conducting disease and preoperative aortic regurgitation.This event is, therefore, a known, inherent risk of the procedure, and it is listed among the potential adverse events in the perceval ifu.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEVAL SUTURELESS AORTIC HEART VALVE
Type of Device
TISSUE HEART VALVE
Manufacturer (Section D)
LIVANOVA CANADA CORP.
5005 north fraser way
burnaby, bc
MDR Report Key9391995
MDR Text Key176228637
Report Number1718850-2019-01190
Device Sequence Number1
Product Code LWR
UDI-Device Identifier00896208000436
UDI-Public(01)00896208000436(240)ICV1210(17)220131
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Type of Report Initial,Followup
Report Date 10/30/2019,02/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/31/2022
Device Model NumberPVS25
Device Catalogue NumberICV1210
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/30/2019
Device Age9 MO
Event Location Hospital
Date Report to Manufacturer10/30/2019
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/29/2019
Supplement Dates Manufacturer Received10/30/2019
Supplement Dates FDA Received02/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Disability;
Patient Age61 YR
Patient Weight101
-
-