SMITH & NEPHEW, INC. GII PS HI FLEX ISRT TR S3-4 9; TRAY, SURGICAL, INSTRUMENT Back to Search Results
Catalog Number 71430408
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/05/2019
Event Type  malfunction  
Event Description
It was reported that the insert trial broke inside the patient during surgery.All the pieces were recovered.There was an s&n backup available and there were no patient injuries.
 
Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device fractured in half and both pieces were returned.The device has numerous nicks and gouges in the surface, particularly around the fracture site.The device was manufactured in 2011 and exhibits signs of extensive wear / usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS HI FLEX ISRT TR S3-4 9
Type of Device
TRAY, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9392041
MDR Text Key168679277
Report Number1020279-2019-04205
Device Sequence Number1
Product Code FSM
UDI-Device Identifier03596010493033
UDI-Public03596010493033
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 12/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71430408
Device Lot Number11KM03355
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2019
Date Manufacturer Received12/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-