MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS DIVISION CELL-DYN RUBY ANALYZER; AUTOMATED HEMATOLOGY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS DIVISION CELL-DYN RUBY ANALYZER; AUTOMATED HEMATOLOGY ANALYZER Back to Search Results
Model Number 08H67-01
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Patient Involvement (2645)
Event Date 09/30/2019
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A final report will be submitted when the evaluation is complete.
 
Event Description
Evidence of fire on fan of the cell-dyn ruby analyzer was observed.No smoke or sparks were observed.No injuries occurred.
 
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling and instrument service.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.A loose wire pin at pin 4 (+15.5vdc) of j18 connector on the interface cable was found and may have caused a short circuit to the adjacent wire at pin 3 (gnd) which had the connectors melt due to an excessive power dissipation.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN RUBY ANALYZER
Type of Device
AUTOMATED HEMATOLOGY ANALYZER
Manufacturer (Section D)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
MDR Report Key9392095
MDR Text Key168526651
Report Number2919069-2019-00114
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier00380740017170
UDI-Public00380740017170
Combination Product (y/n)N
PMA/PMN Number
K061667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 02/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number08H67-01
Device Catalogue Number08H67-01
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/15/2019
Initial Date FDA Received11/29/2019
Supplement Dates Manufacturer Received02/27/2020
Supplement Dates FDA Received02/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-