MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/02/2019
Event Type  malfunction  
Manufacturer Narrative
The device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A visual and microscopic examination of the balloon material identified no issues which could potentially have contributed to this complaint.The returned device was attached to a boston scientific encore inflation unit.Positive pressure was applied but the balloon could not be inflated due to a shaft leak.The markerbands and tip section of the device were visually and microscopically examined, and no issues were noted with the device that may have potentially contributed to the complaint incident.The returned device was attached to a boston scientific encore inflation unit.Positive pressure was applied, and liquid was observed to be exiting from a shaft leak located approximately 40mm proximal to the proximal balloon bond.A microscopic examination found the shaft to have a cut/abrasion on it.This type of damage is consistent with the device coming in to contact with a sharp object.A visual and tactile examination identified that the hypotube was kinked at more than one location.This type of damage is consistent with excessive force being applied to the device.No other issues were identified during the product analysis.
 
Event Description
It was reported that a balloon hole occurred.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified left anterior descending artery.During the procedure, it was noted that a 10mmx3.00mm wolverine coronary cutting balloon had a hole on the balloon on the distal shaft part.A 2.75 x 10 wolverine coronary cutting balloon had the same issue.The procedure was completed with a different device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9396220
MDR Text Key174402811
Report Number2134265-2019-14235
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/05/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-