Model Number IPN046311
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/01/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that the piston of the syringe is smaller than the syringe and the liquid passes between the piston and the inner wall of the syringe.
 
Manufacturer Narrative
(b)(4).A device history record review was performed on the lor syringe with no relevant findings.A design history review was performed for part # kz-05501-002 as a part of this complaint investigation.Per eco-051699 (released 03-dec-2018), supplier (preox) made the following changes: kz-05501-002 luer plastic lor syringe: changed plunger material from profax 535 to profax 531 changed to new plunger tool changed to new mold for blue stopper changed molding location for the plunger and the blue stopper as follows.Plunger: from fleimaplastic in germany to gpe, germany blue stopper: from et, germany to psilkon, germany these effective changes did impact product design and material.A corrective action is not required at this time as the root cause for this complaint investigation could not be determined based upon the information provided and without the actual sample.Complaint verification testing could not be performed as no sample was provided for analysis.A device history record review was performed on the lor syringe with no evidence to suggest a manufacturing related cause.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without the sample.
 
Event Description
It was reported that the piston of the syringe is smaller than the syringe and the liquid passes between the piston and the inner wall of the syringe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION SET
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key9397042
MDR Text Key168688525
Report Number3006425876-2019-01020
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
PMA/PMN Number
K103658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 11/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/01/2021
Device Model NumberIPN046311
Device Catalogue NumberJC-05400-B
Device Lot Number71F19J0768
Was Device Available for Evaluation? Yes
Date Manufacturer Received01/07/2020
Is This a Reprocessed and Reused Single-Use Device? No