SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPAION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 11/13/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the companion 2 driver was not supporting a patient.The companion 2 driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The companion 2 driver was not supporting a patient.The customer, a syncardia authorized distributor, reported that the companion 2 driver did not recognize the two companion external batteries or ac power supply.
 
Manufacturer Narrative
The customer-reported issue of the driver not detecting external batteries or external power was not able to be reproduced during investigation testing.The investigation found that the driver functioned as intended and passed all requirements which included tests to confirm that the driver display showed a plug icon (indicating connection to external main a/c power) and for the hospital cart and driver to both display a plug icon indicating connection to a/c power when the driver was docked to the hospital cart.Additionally, external batteries that have previously met performance requirements were used to test the ability of the driver to recognize and charge external batteries, which the driver was able to do.The external batteries used at the time of the customer-reported issue were not returned and therefore could not be evaluated as part of this investigation.The root cause of the customer-reported issue cannot be conclusively determined.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPAION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key9398601
MDR Text Key168774647
Report Number3003761017-2019-00347
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 12/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2019
Initial Date Manufacturer Received 11/13/2019
Initial Date FDA Received12/02/2019
Supplement Dates Manufacturer Received11/13/2019
Supplement Dates FDA Received01/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-