Model Number 397002-001 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/13/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to a patient because the issue was observed when the companion 2 driver was not supporting a patient.The companion 2 driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The companion 2 driver was not supporting a patient.The customer, a syncardia authorized distributor, reported that the companion 2 driver did not recognize the two companion external batteries or ac power supply.
|
|
Manufacturer Narrative
|
The customer-reported issue of the driver not detecting external batteries or external power was not able to be reproduced during investigation testing.The investigation found that the driver functioned as intended and passed all requirements which included tests to confirm that the driver display showed a plug icon (indicating connection to external main a/c power) and for the hospital cart and driver to both display a plug icon indicating connection to a/c power when the driver was docked to the hospital cart.Additionally, external batteries that have previously met performance requirements were used to test the ability of the driver to recognize and charge external batteries, which the driver was able to do.The external batteries used at the time of the customer-reported issue were not returned and therefore could not be evaluated as part of this investigation.The root cause of the customer-reported issue cannot be conclusively determined.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
|
|
Search Alerts/Recalls
|