(b)(4) reviewed a sales record and confirmed that the subject device was installed to the user facility in 2012.In addition, (b)(4) reviewed the service history of the subject device and confirmed that there was no maintenance performed by (b)(4) until the event occurred.The subject device in this report has not been returned to olympus medical systems corp.(omsc) for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
|
This supplemental report is being submitted to provide additional information.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.However, eight years have passed since the subject device was installed to the user facility, and maintenance was not performed by olympus during the period of use.Therefore, it is surmised that the reported event occurred because the subject device was used in a condition that the insertion section of the subject device was significantly deteriorated.The instruction manual of the device has already warned as follows; the probability of failure of the endoscope and ancillary equipment increase as the number of procedures performed and/or the total operating hours increase.In addition to the inspection before each procedure, the person in charge of medical equipment maintenance in each hospital should inspect the items specified in this manual periodically.An endoscope with which an irregularity is suspected should not be used, but should be inspected by following section 5.1, ¿troubleshooting guide¿ on page 52.If the irregularity is still suspected after inspection, contact olympus.
|