The customer contacted a siemens customer care center (ccc) and reported that discordant hemoglobin (hgb), hematocrit (hct), red blood cells (rbc), and red cell distribution width (rdw) results were obtained on patient samples on an advia 560 hematology system.Siemens is investigating the issue.
|
Discordant hemoglobin (hgb), hematocrit (hct), red blood cells (rbc), and red cell distribution width (rdw) results were obtained on patient samples on an advia 560 hematology system compared to results generated on the lab's advia 2120i hematology system.The discordant results were not reported to the physician(s).The hgb, hct, rbc, and rdw results obtained on the lab's advia 2120i hematology system were considered correct.There are no reports of patient intervention or adverse health consequences due to the discordant hemoglobin (hgb), hematocrit (hct), red blood cells (rbc), and red cell distribution width (rdw) results.
|
Siemens filed the initial mdr 2432235-2019-00442 on 04-dec-2019.Additional information (06-jan-2020): a siemens customer service engineer (cse) was dispatched to the customer site.The cse replaced the system amplifier board as well as the red blood cell (rbc) measure chamber.After the parts were installed, the cse checked the system electronics and pneumatics, both of which were acceptable.The probable cause of the event was a sample specific issue; however, siemens cannot rule out the two replaced parts as a possible cause of the event.The cause of the event is unknown.The system is performing according to specifications.No further evaluation of this device is required.
|