The visual and functional inspection prior to the repair showed that the chuck was worn out, hence it was not able to hold the tool (bur) in position and the running performance was low.This is the result of normal wear process.To avoid such issues the user instruction contains already several notes, warnings and requests how to prepare the handpiece for each treatment and how to use it: warning: hazards for the care provider and the patient.In the case of damage, irregular running noise, excessive vibration, untypical warming or when the cutter or grinder cannot be held.Do not use further and notify service.2.2 technical condition: a damaged device or components could injure patients, users and third parties.Only operate devices or components if they are undamaged on the outside.Check that the device is working properly and is in satisfactory condition before each use.If there are any sites of breakage or clearly recognised changes on the surface, the parts must be checked by the service.Safety checks may only be performed by trained service personnel.If the following defects occur, stop working and have the service personnel carry out repair work: malfunctions.Damage.Irregular noise.Excessive vibration.Untypical heating.No firm seating of the grinder or cutter in the instrument.Observe the following instructions in order to guarantee optimum functioning and prevent material damage: service the medical device with care products and systems regularly as described in the instructions for use.The device should be cleaned, serviced and stored in a dry location, according to instructions, if it will not be used for a longer period.Kavo recommends specifying in-house service intervals where the medical device is brought to a professional shop for cleaning, servicing and a function check.Define the service interval depending on the frequency of use.(b)(4).
|
During a direct resin restoration to tooth #18 the bur separated from instrument, fell into patient's mouth and was swallowed by him.This did not cause any injury to the patient, just a 'scratchy throat' according to dentist.Patient went to emergency room of hospital to have a chest and abdominal, lateral neck x-ray in order to ensure that bur was swallowed and not aspirated.The doctor of the emergency room detected the bur in the stomach of the patient and informed that it should pass without incident.Patient was followed up the following week without any complaints or discomfort.
|