ETHICON ENDO-SURGERY, LLC. ENSEAL G2 ART CURV SEALER 35CM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
Catalog Number NSLG2C35A |
Device Problems
Break (1069); Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/02/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Batch #:t93y7l.Investigation summary: the device was returned with the jaws damaged at the welding point.Upon inspection under magnification of the jaw it was noted that one of the retention pins that holds the jaws in position was sheared off.The device was connected to the generator and it was recognized.Because the jaw was damaged not all functional testing could be performed with the generator.The noises heard could be the retention pin getting broken.The condition of the jaw prevented the functionality of the device.The instrument was unable to fully cycle open and close.A probable cause of the damage to the retention pins could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
|
Event Description
|
It was reported that during a lap hysterectomy, the doctor was about three fourths done with the procedure using the device heard a pop and the jaw broke.They checked the device and it looked like it was all there.Nothing fell in the patient.The case was extended by a few minutes.Case completed with another device of the same product code.There were no patient consequences reported.
|
|