• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENSEAL G2 ART CURV SEALER 35CM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENSEAL G2 ART CURV SEALER 35CM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number NSLG2C35A
Device Problems Break (1069); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/02/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch #:t93y7l.Investigation summary: the device was returned with the jaws damaged at the welding point.Upon inspection under magnification of the jaw it was noted that one of the retention pins that holds the jaws in position was sheared off.The device was connected to the generator and it was recognized.Because the jaw was damaged not all functional testing could be performed with the generator.The noises heard could be the retention pin getting broken.The condition of the jaw prevented the functionality of the device.The instrument was unable to fully cycle open and close.A probable cause of the damage to the retention pins could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Event Description
It was reported that during a lap hysterectomy, the doctor was about three fourths done with the procedure using the device heard a pop and the jaw broke.They checked the device and it looked like it was all there.Nothing fell in the patient.The case was extended by a few minutes.Case completed with another device of the same product code.There were no patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL G2 ART CURV SEALER 35CM
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
6107428552
MDR Report Key9416696
MDR Text Key205649837
Report Number3005075853-2019-23992
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10705036014508
UDI-Public10705036014508
Combination Product (y/n)N
PMA/PMN Number
K122797
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberNSLG2C35A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2019
Date Manufacturer Received11/06/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/26/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-